Connection using the handwheel
• Make absolutely sure that the groove inside the gasket is not damaged when
making the handwheel connection.
If necessary, wipe the groove for the gasket clean with a dry cloth. You can also
moisten the cloth with a little clean water. The safety and function of the appliance
can only be assured if original TESCOM EUROPE spare parts are used.
Carefully push the new gasket into the connection groove. Do not use a tool to do
this under any circumstances.
Connection with pin index
• Make absolutely sure that the groove inside the gasket is not damaged.
If necessary, wipe the sealing surface clean with a dry cloth. You can also moisten
the cloth with a little clean water.
The safety and function of the appliance can only be assured if original TESCOM
EUROPE spare parts are used.
Carefully push the new gasket onto the connection sealing surface. Do not use a tool
to do this under any circumstances.
Connection by means of a threaded connection
• Make absolutely sure that there is no damage to the groove or the connection
piece sealing surface in which or on which the gasket is located.
If necessary, wipe the groove for the gasket clean with a dry cloth. You can also
moisten the cloth with a little clean water. The safety and function of the appliance
can only be assured if original TESCOM EUROPE spare parts are used.
Carefully push the new gasket into the connection groove.
Flat sealing discs must be placed on the sealing surface of the connection piece for
the pressure regulator. Do not use a tool to do this under any circumstances.
6.3

Maintenance and service life

The service life of the appliance is 12 years from the month and year of manufacture,
which are stated on the type plate.
61