DESCRIPTION
Battery Shut Down
Attention!
If battery shutdown alarm activates, stop using the instrument
as it no longer has enough power to indicate potential hazards,
and persons relying on this product for their safety could sustain
serious personal injury or death.
The instrument goes into battery shutdown mode 60 seconds before final
shutdown (when the batteries can no longer operate the instrument):
"
Batt Alarm
Alarm sounds and lights flash; alarm cannot be silenced
No other pages can be viewed; after approximately one minute, the
instrument automatically turns off.
When battery shutdown condition occurs:
(1) Leave the area immediately.
(2) Recharge the battery.
Battery Charging
Attention!
Risk of explosion: Do not recharge instrument in hazardous
area.
Caution!
Use of any charger, other than the Charger supplied with the
instrument, may damage or improperly charge the batteries.
The charger is capable of charging a completely depleted pack in
less than four hours in normal, room-temperature environments.
Allow very hot or cold instruments to stabilize for one hour at
room temperature before attempting to charge.
Minimum and maximum ambient temperature to charge the
instrument is 10°C (50°F) and 35°C (95°F), respectively.
For best results, charge the instrument at room temperature
(23°C).
14
" and
ALTAIR 4 - Operating Manual 10088160/00 Y-HLR
flash on the display
MSA
GB