Descargar Imprimir esta página

Agilent Technologies IPCMini Manual De Usuario página 195

Publicidad

Technical Information
Procedure to Connect the Serial and I/O ports
to an External Cable
In the following picture, it has showed the right procedure to connect a cable to the
I/O or to the serial port. A shielded cable of 30 m maximum length has to be utilized
for both serial and I/O port connection.
Take care to have a good contact between the metallic case and the external shield
NOTE
of the wire. Moreover, this operation should be done at least on the controller side
The I/O and communication terminal provided at the equipment are only required to
be connected with external circuits which are separated from hazardous live
voltages by reinforced or double insulation.
In this way, you will be sure to reduce the influence of the external noise and to
accomplish the EMC requests. In picture d is showed the cable assembled.
Figure 62
Agilent IPCMini Ion Pump Controller User Manual
195

Publicidad

loading