Amprobe MEGATEST 1000 Manual Del Propietário página 17

Publicidad

Idiomas disponibles

Idiomas disponibles

5.2.2.1 "MAN" MODE
GO
At the end of the test, in case the
resistance
results to be lower than R
(depending on the selected voltage
see following Table 4) and the test is
effected at the selected rated
voltage value, the instrument emits a
double sound signal indicating the
positive
outcome
displays one screen similar to the
screen alongside.
The values of measured resistance of isolation always must be confronted with the
normative limits (see Errore. L'origine riferimento non è stata trovata.) for being able to
assert if the system is to norm.
SAVE
Press the GO key.
The instrument effects the test lasting:
Maximum 10 seconds in case the key is pressed and released
within 5 seconds.
Until the key is released for all the other cases.
Note: Press the GO key again if the test is to be stopped
value
MΩ
detected
of
the
Rated voltage value selected for the test.
The test can be stored pressing the SAVE key twice (according to
paragraph 0).
R
1.00
MAX
500 V 20
MAN
test
16
Insulation
resistance
ISO
value R
MΩ
Test duration.
s
In this case the GO
key
pressed
seconds.
Manual mode.
ISO.
has
been
for
20