Example:
97 tests have been stored in the instrument.
If you want to cancel the tests from 43rd to 97
-
Press RCL.
-
Select with keys
-
Press CLR. You can see CLR blinking on the primary display, 43 on the
left secondary display and 97 on the righ secondary display.
Press CLR. The tests from 43 to 97 has cancelled.
-
5.4
RESET PROCEDURE
Reset of the INSTRUMENT and default parameters
Press DISP, CLR, RCL at the same time the rotary switch is turned on.
The screen nearby is displayed
for 5 seconds, after which the
instrument emits a sound signal
and then displays the screen
relative to the function selected
with the rotary switch.
ATTENTION: the reset procedure will erase all the stored tests and it will set the default
parameters in the instrument.
Before conducting a reset procedure, download the stored tests on a PC.
5.5
DEFAULT PARAMETERS
5.5.1 DEFAULT PARAMETERS LOW
Parameter
Mode
Calibration Offset
Mode R+/R- TIMER
5.5.2 DEFAULT PARAMETERS M
Parameter
Mode
Test voltage
Mode TIMER
5.5.3 MEMORY DEFAULT PARAMETERS
Parameter
Parameter P
Memory state
,
the memory location 43.
Reset default parameter
Reset default parameter
Reset default parameter
24
th
, you proceed as follows:
rES
AUTO
0 (CAL off)
Timer 3s
MAN
500V
Timer 60s
P = 1
1(empty)