10. TECHNICAL SPECIFICATIONS
TECHNICAL FEATURES
Accuracy is indicated as [% of reading + number of digits]. It refers to the following
atmospheric conditions: a temperature of 23°C ± 5°C with a relative humidity < 75%.
Continuity
Test mode
AUTO, R+, R-
R+TIMER
R-TIMER
Test current
Test current resolution
Open circuit Test Voltage
Insulation resistance
Test voltage
1000
Automatic selection of the measuring range
Open circuit voltage
Short circuit current
Rated measuring current
If the instrument measures a voltage higher than 30V, the instrument will not take a
measurement.
Safety standards
Insulation:
Over voltage category:
Measuring range
(Ω)
0.01 – 19.99
20.0 – 99.9
0.01 – 9.99
>200mA DC up to 16Ω (incluse the resistance of the calibration)
40mA DC from 16Ω to 99.9Ω
Measuring range
(Ω)
0.01 – 19.99
50
20.0 – 49.9
50.0 – 99.9
0.01 – 19.99
20.0 – 99.9
100
100.0 – 199.9
0.01 – 19.99
20.0 – 199.9
250
200 - 249
250 - 499
0.01 – 19.99
20.0 – 199.9
500
200 - 499
500 - 999
0.01 – 19.99
20.0 – 199.9
200 - 999
1000 - 1999
Resolution
(Ω)
0.01
0.1
0.01
1mA
9V
Resolution
(MΩ)
(Ω)
0.01
0.1
0.1
0.01
0.1
0.1
0.01
0.1
1
1
0.01
0.1
1
1
0.01
0.1
1
1
1,1 x Rated voltage test
<3,0mA - 500V
<2,0mA - 50V, 100V, 250V, 1000V
2,17 mA - 500V/230kΩ
1mA - 1000V/1MΩ
Class 2, double insulation
CAT III
37
Accuracy
±(2% Reading + 2 digit)
±(2% Reading + 2 digit)
Accuracy
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)
±(2% Reading + 2 digit)
±(5% Reading + 2 digit)