08.4 Û CLEANING AND STERILIZATION OF THE INSERTS
PREPARATION
- Disconnect the insert from the handpiece
with the torque wrench (see Chapter 03)
MATERIALS NECESSARY
- Enzymatic detergent with pH 6-9
- Water
- Ultrasound tank
- Brush with soft nylon bristles
- Compressed air
- Distilled water
- Single-use sterilization packets
- Steam sterilization
- Syringe
CLEANING AND DISINFECTION METHOD - Inserts
10 min
08 Û CLEANING AND STERILIZATION
from the handpiece before proceeding to
its cleaning and sterilization.
peroxide. If you intend to disinfect the
inserts, do not use hydrogen peroxide, but
only disinfectants with neutral pH (pH7);
always rinse with sterile water to preserve
the disinfection.
of the enzymatic detergent solution, do
not recycle.
Prepare an enzymatic detergent solution
1
with pH 6-9, as per the instructions of the
manufacturer;
Place the insert in the ultrasound tank
2
submerged by the enzymatic detergent
40°C
solution at 40°C, for at least 10 minutes;
CAUTION: Always disconnect the insert
CAUTION: Do not use hydrogen
CAUTION: Once used, correctly dispose
EN
29