05.3 Û SAFETY PRECAUTIONS BEFORE AND DURING USE
WARNING: Before starting to operate
with the device, make sure that you
have stock material (handpiece, inserts,
wrenches) available to use in case
EN
the device stops due to a fault or of
inconveniences.
WARNING: Only use original Mectron
inserts, accessories, and spare parts.
WARNING: Use of non-original
Mectron inserts: this use entails finite
damage to the handpiece threading,
thus compromising correct operation and
risking to cause harm to the patient.
CAUTION: Contraindication. Ultrasonic
scaler. Do not perform treatments on
prosthetic artifacts made of metal or
ceramics. The ultrasonic vibrations could
lead to the de-cementing of the artifacts.
WARNING: Contraindications. Do
not use the multipiezo on patients who
carry heart stimulators (Pace-makers) or
other implantable electronic devices. This
precaution also applies to the operator.
WARNING: Contraindications. Do not
perform scaling treatments without water
spray in order to avoid the insert over-
heating which may cause damages to the
tooth. Treatments with no water spray can
be carried out only with "Dry Work" inserts
which do not have the water passage.
CAUTION: Only use inserts with water
passage when performing treatments that
require irrigation.
WARNING: Treatments that require
irrigation. Always check operation of the
irrigation before and during use. Make sure
the fluid outflows from the insert.
Do not use the device if the irrigation does
not work or if the pump is defective.
WARNING: Checking device status
before the treatment. Always check that
there is no water underneath the device.
Before every treatment, always check that
the device works perfectly and that
16
the accessories are efficient. In case you
uncover operating abnormalities, do
not perform the treatment. Contact an
Authorized Mectron Service Center if the
abnormalities concern the device.
WARNING: Infections control.
First use: All new and repaired accessories
are supplied in NON STERILE conditions.
Before use, and after each treatment,
they must be cleaned and sterilised in
strict compliance with the instructions
given in Chapter 08 - CLEANING AND
STERILIZATION.
Subsequent uses: After every treatment,
clean and sterilize all the reusable parts
and accessories, following the instructions
provided in Chapter 08 - CLEANING AND
STERILIZATION.
WARNING: To ensure that the
handpiece cools down, always activate it
with the irrigation circuit correctly installed
and filled. To fill the irrigation circuit,
always use the FLUSH function.
CAUTION: To correctly use the device,
you must press the foot pedal and start
it with the insert not making contact
with the part to be treated, so that the
electronic circuit can recognize the best
point of resonance of the insert without
interferences, allowing its optimal
performance.
WARNING: Before every treatment,
make sure that the insert appropriate for
the treatment is inserted on the handpiece.
Exclusively use the Mectron torque wrench
to fasten the insert to the handpiece.
WARNING: The patient must not come
into contact with the device body or the
foot pedal.
WARNING: Do not change the insert
while the handpiece is operating, to
prevent causing wounds to the operator.
CAUTION: FLUSH function. After the
device is used with aggressive and non-
aggressive solutions, it is necessary to