MAINTENANCE AND USE MANUAL
Should any of the anomalies described below occur try setting the situation right, as
far as is permissible, by following the instructions contained in this manual. Should
the problem persist contact your Authorised Servicing Centre.
p
roblem
It is difficult to
create
a
vacuum in the
containers.
After turning the
main switch to
ON the machine
does not work.
The pump does
not start.
- Error n°1: if, 2 minutes after positioning the cup, the pressure still hasn't dropped below the
threshold (= 50) a warning appears and the pump is switched off. In this case it is necessary to carry
out an inspection and trace the problem (e.g. the cup may be positioned incorrectly, the vacuum
sensor tube disconnected, the pump disconnected...) before restarting the cycle.
- Error n°2: if, during the cycle, the cup is no longer detected, a warning appears and the
outputs are deactivated. In this case it is necessary to carry out an inspection and trace the
problem and then restart the cycle. This error may occur if the vacuum sensor is faulty:
measured pressure may be below the threshold (otherwise error 3 would be shown) but the cup
detaches. In this case the motor is stopped to prevent injury to the operator.
- Error n°3: if, during the cycle, the vacuum is insufficient (i.e. pressure rises above the threshold = 50)
a warning appears and the motor is switched off. this case it is necessary to carry out an inspection
and trace the problem (e.g. cup detached, vacuum sensor tube disconnected, pump disconnected...)
before restarting the cycle.
- Error n°4: when the elite mix control unit is switched on the pump is activated for 5 seconds to
check whether the filter is dirty (i.e. without the cup): if the pressure drops below a certain level this
means that the filter is clogged and an error warning appears. In this case the filter must be replaced.
ELite mix
EN
CHAPTER 9: TROUBLESHOOTING
C
ause
1. The gaskets between the
various components are dirty,
damaged
or
positioned
incorrectly.
2. Faulty pump.
3. Intake filter clogged.
1. Plug is not connected to the
mains socket.
Power lead is damaged.
2. Fuse blown.
1. Sensor on board and/or
magnet on cup lid faulty.

PROBLEMS, CAUSES AND REMEDIES

1. Examine the gaskets, clean if necessary
and lubricate them with Vaseline. Clean
mixing unit components thoroughly (see
paragraph 8.1.1).
2. Contact an Authorised Service Centre.
3. Change the filter (see section 8.1.2).
1. Connect the plug to the mains socket
properly.
2. Change the fuse, using one of identical
characteristics. Contact an Authorised
Servicing Centre only.
1. Contact an Authorised Service Centre.
ERROR WARNINGS
WARNING
9.1
9.2
r
emeDy
9.3
127 of 264

Publicidad