Figure 9� Reference Only
The Console will automatically pull saline into the pump.
6.
Place the foot switch within easy access of the physician. Choose a location
that will minimize accidental activation.
Prime the Catheter
The status panel displays PRIME indicating that the initial pump prime was successful.
The time display shows the time required to complete the priming sequence
(Figure 10).
Figure 10� Reference Only
Black (K) ∆E ≤5.0
/ CMYK Image
Catheter priming time is determined by the catheter model and is automatically set
by the Console.
1.
Prime the catheter by completely submerging the tip in heparinized saline and
pressing the foot switch (Figures 11 and 12).
Figure 11�
2.
Continue priming until the time display reaches zero seconds.
The status panel displays PRIME COMPLETE (Figure 13).
Figure 13� Reference Only
3.
Confirm AngioJet™ System set-up is successfully completed by removing foot
from foot switch.
The status panel displays READY and green icon is illuminated (Figure 14).
Figure 14� Reference Only
8
Figure 12�