10. Specifications

10. Specifications
Sensor technology
EN
Measuring range
Overload safety
Sensor
Indicator
Accuracy (density
measurement)
Accuracy (pressure
measurement)
Compensated
temperature range
Base instrument
Units
Density units SF₆
Pressure units SF₆ at
20 °C
Pressure units
Indicator
Display
Rotatable case
Resolution
Functions
Communication
Measuring rate
Menu languages
Memory
38
See product label
3 times
> 110 % FS or -10 % FS
0.6 % FS
0.05 % FS
-10 ... +50 °C [14 ... 122 °F]
g/litre, kg/m³
bar abs., psi abs., KPa abs., MPa abs.
bar rel., psi rel., KPa rel., MPa rel.
Standard: psi, bar
Can be set to: mbar, kg/cm², Pa, hPa, kPa, Mpa, mmH₂O, mH₂O,
inH₂O, inH₂O (4 °C),
inH₂O (20 °C), inH₂O (60 °F), ftH₂O, mmHg, cmHg, inHg, inHg
(0 °C), inHg (60 °F), kp/cm², lbf/ft², kN/m², atm, Torr, micron, as well
as user-defined units
5 ½-digit 7-segment display (incl. a large matrix area for auxiliary
information)
Bar graph, 0 ... 100 %
Selectable backlighting
The case is rotatable by 330°.
4 ... 5 ½ digits; adjustable; depending on the selected pressure unit
Integrated Bluetooth interface, configuration, data transfer and
export with the WIKA "myWIKA device" app (available free-of-
charge in Google Play Store and Apple App Store)
Max. 3/s (density measurement)
Max. 50/s (pressure measurement)
English, German, Italian, French, Spanish, Russian and Polish
Integrated data logger
WIKA operating instructions, model GDI-100-D