Specifications
Physical Characteristics
Dimensions
14�5" high x 10�5" deep x 11" wide
(36�2 cm high x 26�7 cm deep x 28
cm wide)
Weight
9�93 lb (4�5 kg)
Relative noise level
50 dBA
Hose
Flexible
Length: 80" (203 cm)
Weight: 1�51 lb (0�686 kg)
Diameter: 2" (5�08 cm)
Unit Intake Filter
MERV 14*
Recommended Filter
Every 12 months or 500 hours of use�
Change
* See Frequently Asked Questions on Filtration
Recommended operating environment
Temperature
15°C-25°C (59°-77°F)
Humidity
Up to 80% relative humidity (non-
condensing)
Altitude
Max 2,000 m
Operating temperatures
Temperature control
Electronically controlled
Heat generated
1600 BTU/hr (average), 470 W
(average)
Time to reach operating
2 - 5 minutes (dependent on blanket
temperature
model)
Time required for the contact surface
temperature to heat up from 23 ±
2°C to 37°C (73 ± 2°F to 99°F)
Storage/transport
-20°C to 45°C (-4°F to 113°F)
temperature
Store all components in a cool, dry
place when not in use�
10
Safety System
Thermostat
Independent electronic
circuit; thermal cutoff shuts
the heater OFF to ensure
hose end air remains below
56°C (53°C ± 3°C typical)�
Alarm system
Over-temperature (≤56°C,
53°C ± 3°C typical): red
Over-Temp indicator symbol
will flash alternately with
the "Wrench" symbol,
alarm sounds, heater and
blower shut down, operating
indicator lights turn OFF,
and control panel becomes
unresponsive� Remove from
service�
Under-temperature (<3°C
of selected temperature/
setting for 3 minutes):
blue Under-Temp indicator
symbol will flash alternately
with the "Wrench" symbol,
alarm sounds, heater and
blower shut down, operating
indicator lights turn OFF,
and control panel becomes
unresponsive� Remove from
service�
Fault
"Wrench" symbol displays
and alarm sounds� Heater and
blower shut down, operating
indicator light turns OFF
and control panel becomes
unresponsive� Service
required�
Overcurrent protection
Dual input fused lines� Fuses
are to be accessed and/
or replaced only by service
personnel�
Alarm/Fault Codes
Alarm conditions
General fault
Over-temperature
Under-temperature
Fault code indicator
Yes on LCD screen
Filter change indicator
Yes: Triggered by 500 hours
of use
Electrical Characteristics
Heating element
1190 W Resistive
Heater control
3 stage heater
flicker free
Temperature settings
Ambient
32° C (89�6° F)
38° C (100�4° F)
43° C (109�4° F)
Temperature accuracy
+/- 3° C
Temperature display
0�1° C
resolution
Leakage current
Meets IEC 60601-1 requirements
Blower motor
Non-adjustable, one speed
Airflow: up to 44 cfm or 20�77 L/s
Operating speed: 4,700 RPM
Power consumption
Peak: 1255 W
Average: 650 W
Guidance and manufacturer's declaration – electromagnetic emissions
The model 675 is intended for use in the electromagnetic environment specified below� The customer or the user of the model 675 should
assure that it is used in such an environment�
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions,
Group 1
The model 675 uses RF energy only for its internal function� Therefore, its RF emissions are very low
CISPR 11
and are not likely to cause any interference in nearby electronic equipment�
RF emissions,
Class A
NOTE: The EMISSIONS characteristics of this equipment make it suitable for use in industrial areas
CISPR 11
and hospitals (CISPR 11 class A)� If it is used in a residential environment (for which CISPR 11 class
B is normally required) this equipment might not offer adequate protection to radio-frequency
Harmonic
Class A
communication services� The user might need to take mitigation measures, such as relocating or
emissions, IEC
re-orienting the equipment�
61000-3-2
Voltage
Complies
fluctuations/
Flicker emissions,
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The model 675 is intended for use in the electromagnetic environment specified below� The customer or the user of the model 675 should
assure that it is used in such an environment�
These guidelines may not apply in all situations� Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people�
IEC 60601
Electromagnetic environment - guidance
test level, Compliance level
Electrostatic discharge (ESD), IEC 61000-4-2
±8 kV contact
Floors should be wood, concrete or ceramic tile� If floors are covered with synthetic material, the
±15 kV air
relative humidity should be at least 30%�
Electrical fast transient/burst, IEC 61000-4-4
±2 kV for power supply lines
Mains power quality should be that of a typical commercial or hospital environment�
±1 kV for input/output lines
Surge, IEC 61000-4-5
±1 kV line to line
Mains power quality should be that of a typical commercial or hospital environment�
±2 kV line(s) to earth
Voltage dips, short interruptions and voltage variations on power supply input lines, IEC 61000-4-11
Power cord
15 ft�, SJT, 3 cond�, 13 A
15 ft�, SJT, 3 cond�, 15 A
4�6 m, HAR, 3 cond�, 10 A
Device ratings
110-120 VAC, 50/60 Hz, 11 A, or
220-240 VAC, 50/60 Hz, 6 A, or
100 VAC, 50/60 Hz, 12 A
Fuses
Type: Fast acting ceramic fuses, 250
VAC AMP rating:
12 A (110 - 120 VAC)
8 A (220 - 240 VAC)
15 A (100 VAC)
Operating speed: F (Fast Acting)
Breaking capacity:
15 A, 12 A: 750 A @ 250 Vac
15 A, 12 A: 10,000 A @ 125 Vac
8 A: 200 A @ 250 Vac
8 A: 10,000 A @ 125 Vac