Preventative Maintenance; Nurse Call Battery; Main Bed Power Circuit Breaker - Stryker EPIC II 2031 Manual Del Usuario

Cama de cuidados críticos
Tabla de contenido

Publicidad

WARNING
Service only by qualified personnel. Refer to the maintenance manual. Verify the power cord is unplugged before
servicing.

Nurse Call Battery

• To prevent a low battery condition when the bed is not plugged in, position the cord out switch at the head end
of the bed to the off position. The switch is identified by the label shown below. If the switch is not positioned
as shown below and the bed power cord and pendant cord are unplugged, the life of the back-up battery will be
significantly reduced.
• If the power light (located on the foot board) is flashing, the Nurse Call battery needs to be replaced. The battery
is located on the patient's left side at the head end of the bed. No tools are required to replace the battery. Unplug
the bed power cord from the wall socket and replace the battery. Properly dispose of the old battery in accordance
with local regulations.

Main Bed Power Circuit Breaker

In the event of a loss of bed functions, unplug the bed power cord from the wall socket and reset the circuit breaker(s)
located under the bed on the patient's left side. Plug the bed into a properly grounded wall receptacle and follow the
setup procedures listed on
page
www.stryker.com

Preventative Maintenance

2-14.
2031-009-011 REV B
Return To Table of Contents
2-25