CARE & MAINTENANCE
The DeVilbiss iFill Oxygen Cylinders should be kept clean and free from moisture and dust. The cylinders should be protected
from extreme temperatures. Clean the cylinders periodically by wiping with a dry, lint-free cloth. Do not clean with a solvent based
cleaning solution. Avoid dropping the cylinder or placing it in a position where it could topple or fall since this can damage the
device. Whenever possible, use a padded carrying bag such as the Hideaway or Walkabout bags to carry the cylinder. This will help
to protect it in case of a fall. Avoid getting fluids or debris such as sand or dirt inside the oxygen cylinder and/or oxygen connector.
Do not place oxygen cylinders in unventilated spaces such as car trunks. Excessive heat can make the relief valve sud-
denly and quickly discharge the cylinder contents, possibly making it a projectile and greatly increasing the oxygen
Do not leave oxygen cylinders in the cabs of vehicles without ventilation. If a cylinder leaks, a spark could start a
TROUBLESHOOTING
Do not attempt to open the device for maintenance or repair. The device contains no
user-serviceable parts. Contact your oxygen provider if service is required.
Symptoms
Oxygen is not being delivered even
though the PulseDose indicator is flash-
ing every time I breathe.
Use times are different from those stat-
ed in the literature.
PulseDose will not pulse
PulseDose works fine for a couple of
minutes, then sensitivity seems to drift
and may stop working altogether.
STORAGE AND HANDLING WARNINGS
level in unventilated spaces.
fire causing serious injury or death.
Remove cylinders from the vehicle when the destination is reached.
Be sure to secure cylinders from movement during transport.
1. Oxygen supply is empty.
1. PulseDose responds to your breath
rate. Your breath rate may vary,
which causes the operation time to
vary.
2. Leak in system
1. Cannula is not adjusted properly.
2. Unit is not turned on.
3. Batteries discharged or not
installed.
4. Mouth breathing with closed soft
palates.
5. Unit did not reset while changing
batteries (red light stays on).
1. Using pediatric cannula or any can-
nula that restricts continuous flow
capacity of 10 lpm.
C A R E & M A I N T E N A N C E / T R O U B L E S H O O T I N G
WARNING
Possible Causes
11
Remedies
1. Check contents indicator on the
device. If empty, switch cylinders.
1. PulseDose is operating correctly.
2. Contact your Sunrise Medical
provider.
1. Check all cannula connections to
make sure they are tight, and
adjust the cannula to fit comfort-
ably in your nose. Ensure tubing is
not kinked.
2. Turn the rotary selector to the
appropriate setting.
3. Install new batteries.
4. Breathe through the nose (cannula).
5. Turn the unit off and back on using
the rotary selector.
1. Replace with standard nasal cannula.
SE-535-CYL