"The coil temperature is too low"
"The coil temperature is too high"
"A coil current peak was detected,
Attention: the yoke has not been
demagnetized!"
"A coil voltage peak was detected,
Attention: the yoke has not been
demagnetized!"
36
The coil temperature -10°C.
The coil temperature 120°C.
A current peak has been detected.
A voltage peak has been detected
exceeding 500V
Switch off the device and wait until the
ambient temperature has risen above
-10°C (14°F). If the temperature is within
the limit and the error still occurs,
please contact your supplier.
Switch off the device and wait until
the ambient temperature has dropped
below 120°C (248°F). If the temperature
is within the limit and the error still
occurs, please contact your supplier.
Switch off the device and wait a few
seconds before switching it on again.
Switch off the device and wait a few
seconds before switching it on again.