Figure 24 - Load side emergency button
12.5 Replacement of general disjoints
The general circuit breaker is located in the electrical control of the generator.
Disconnect the electrical power from the mains before starting the component replacement.
To replace the general circuit breaker, proceed as follows:
With the sterilizer off, open the side trim using the key provided with the equipment;
Access the generator's electrical control;
Remove the connected cables with the aid of an allen key;
Remove the two screws fixing the circuit breaker with the aid of a philips wrench;
Remove the general circuit breaker from the control;
Reposition the new general circuit breaker and install the fixing screws;
Reposition the input and output cables of the circuit breaker and make the connections again;
Turn on the equipment and test the general circuit breaker to validate its operation.
12.6 Replacement of the bipolar circuit breaker
The bipolar circuit breaker is located in the electrical control of the generator.
Disconnect the electrical power from the mains before starting the component replacement.