Descargar Imprimir esta página

Baumer HI SPEED II B0111-100 Manual De Mantenimiento página 49

Esterilizador a vapor

Publicidad

Idiomas disponibles

Idiomas disponibles

15.4 Fault indications (alarms)
HI SPEED II line sterilizers have an extremely functional control logic so that they do not allow actions or
parameters to be outside the required ranges or conditions, which makes this product extremely reliable during
the sterilization cycle.
The sterilizer alarms may occur under different circumstances or process steps.
The alarms are different from the other parameters because they are flashing intermittently on the screen.
Maintenance Manual
Steam Sterilizer
HI SPEED II
Calibration "PT3".
Internal chamber calibration
Connect the pressure gauge to the transducer
of the internal chamber.
Depressurize the calibrator using the relief
valve of the pressure pump and adjust the
calibrator pressure to 0.400 bar and press the
"ZERO" key.
Insert the 2.500 bar pressure pump into the
pressure gauge and press the "SPAM" button.
Repeat the "ZERO" and "SPAM" settings three
times.
After performing the pressure calibration, select the
icon for "Atmospheric pressure".
Obs.: The loading door must be open to adjust the
atmospheric pressure.
Figure 38 - Fault indication
Code: MM.01.09.0003
Version / Revision: 3.001
Page: 49 of 81

Publicidad

loading