See below the main alarms and their causes and solutions:
Problem
Sterilization Overtemperature
Alarm.
Sterilization Undertemperature
Alarm.
Parameter Range Failed.
Vacuum Pump Thermal Relay
Failure.
Equipment does not have a TTS
Sensor.
Close Discharge Port.
Close Loading Door and then
start the cycle.
Steam Pressure Switch (Program
B) failure.
Air Pressure Switch (Program B)
failure.
Vacuum Pump Water Pressure
Switch (Program B) failure.
Overpressure in the External
Chamber. Disconnect the
sterilizer (Program B).
Internal Chamber Leak Test failed
(Program B).
Maintenance Manual
Steam Sterilizer
HI SPEED II
Cause
 Failure of the steam supply valve to
the inner chamber.
 Defective steam valve for the inner
chamber.
 Vacuum system failure.
 Steam chamber failure in the inner
chamber.
 Vacuum pump motor circuit breaker
turned off.
 Vacuum pump overload.
 Attempted to select the TTS sensor
without the sterilizer having this
option.
 Energized sterilizer with open
discharge door.
 Attempt to start the cycle with the
loading door open.
 Generator control circuit breakers
turned off.
 Failure to supply water to the
generator.
 Burned heating elements.
 Compressed air supply failure.
 Lack of water to feed the vacuum
pump.
 Failure to control the pressure of
the inner chamber.
 Air or steam entering the inner
chamber.
Code: MM.01.09.0003
Version / Revision: 3.001
Page: 50 of 81
Corrective measure
 Check for steam leakage through
the valve of the inner chamber.
 Replacement of the steam valve for
the internal chamber.
 Check the operation of the vacuum
pump and the vacuum valve of the
inner chamber.
 Check the operation of the steam
valve of the inner chamber.
 Reset the vacuum pump motor
circuit breaker.
 Check the electric current of the
vacuum pump.
 Do not select the TTS sensor or
install the option.
 Press the button to close the
discharge door.
 Close the door and then start the
cycle.
 Turn on the generator control circuit
breakers.
 Check the water supply to the
generator.
 Replace resistors.
 Check compressed air supply.
 Check the water supply to the
vacuum pump.
 Turn off the sterilizer.
 Check connections and valves of
the inner chamber.