Cleaning and Maintenance
22
Operating Instructions Quintix
8.2
Dismounting the Draft Shield
Requirements
— The side doors of the draft shield are opened.
— The weighing pan and the associated components have been dismount-
ed.
Procedure
Lift the front of the draft shield base plate and remove.
t
To release the draft shield: Turn the clamping ring in the direction of the
t
marking
.
Remove the draft shield from the device with both hands.
t
8.3
Cleaning the Device
Procedure
Disconnect the power supply cable from the power supply at the instal-
t
lation site.
Only use suitable cleaning agents and cleaning procedures and observe
t
the product information for the cleaning agent used (see Chapter "12.10
Cleaning Agents and Cleaning Procedures", page 29).
Remove dust and powdery sample residue with a brush.
t
Wipe the device and the associated components with a slightly damp
t
cloth. Use a suitable cleaning agent for more severe contaminations (see
Chapter "12.10 Cleaning Agents and Cleaning Procedures", page 29).
8.4
Software Update
Procedure
To perform a software update, contact Sartorius Service.
t
Pro
®