12.11.4 Models QTX12201Ix | QTX10201Ix | QTX8201Ix | QTX6201Ix | QTX3201Ix
Model
Scale interval (d)
Maximum capacity (Max)
Weighing system
Repeatability
At 5% load, typical value
At approx. the maximum load, typi-
cal value
Linearity deviation
Limits
Typical value
Eccentricity deviation: Deviation when load is off-center, positions according to OIML R76
Test weight
Tolerance
Typical value
Sensitivity drift between +10°C and
+30°C
Tare maximum capacity (subtractive)
isoCAL:
Temperature change
Time interval
Only for models with approval:
Accuracy class
Type
Verification scale interval (e)
Minimum load (Min)
Minimum initial weight according to USP (United States Pharmacopeia),
Chap. 41
Optimum minimum initial weight
Typical minimum initial weight
Typical measurement time
Typical stabilization time
Unit
Value
mg
100
g
12,200
EMC
± mg
50
± mg
100
± mg
100
± mg
60
g
5,000
± mg
200
± mg
100
± ppm/K
4
K
2
h
6
II
BC-QG
mg
1,000
mg
5,000
g
82
g
100
s
≤ 1.0
s
≤0.9
Technical Data
Value
Value
Value
100
100
100
10,200
8,200
6,200
EMC
EMC
EMC
50
50
50
100
100
50
100
100
100
60
60
60
5,000
5,000
2,000
200
200
200
100
100
100
4
4
2
2
2
2
6
6
6
II
II
II
BC-QG
BC-QG
BC-QG
1,000
1,000
100
5,000
5,000
5,000
82
82
82
100
100
100
≤ 1.0
≤ 1.0
≤ 1.0
≤0.9
≤0.9
≤0.9
Operating Instructions Quintix
Value
100
3,200
EMC
50
50
100
60
2,000
200
100
2
2
6
II
BC-QE
100
5,000
82
100
≤ 1.0
≤0.9