Cpr Procedure - Invacare SoftCloud Ace Manual Del Usuario

Ocultar thumbs Ver también para SoftCloud Ace:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 111
the user, ideal for the most vulnerable users such as those
with limb ischaemia or existing wounds.
CAUTION!
Insufficient pressure relief
– Deflating more than one adjacent air cell can
affect the support of the user during the
normal alternating support cycle. This function
should only be used for temporary procedures.
A
B
1. Open the push button flap A at the foot end of the
mattress to access the 5 release connectors.
2. Disconnect the chosen connector B for permanent
offloading.
Top connector is closest to the knee.
Connector 4–2 is below the knee and goes
towards the heel.
Bottom connector is the last heel cell
closest to the end of the mattress.
It is not recommended to deflate more than one
air cell directly beneath the area where relief
is needed.
1630732-B
3. Reconnect the connector(s) to disengage permanent
offloading. The air cell will be inflated again and return
to normal operation mode.
4. Close the push button flap A.

5.5 CPR Procedure

A
1. Firmly pull the yellow rapid release CPR tag A from the
side of the mattress to deflate the whole system.
2. Switch off the control unit.
Mattress will start to deflate.
3.
When CPR is complete replace the yellow CPR Tag
ensuring the sealing connectors are firmly attached and
restart the control unit following chapter "Activating the
control unit". Using the target design to line up each
plug with it's corresponding socket.
Usage
19

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Softcloud pro

Tabla de contenido