Invacare SoftCloud Ace Manual Del Usuario página 33

Ocultar thumbs Ver también para SoftCloud Ace:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 111
Immunity test
IEC 60601 test level
3 V
Conducted RF
150 kHz to 80 MHz
IEC 61000-4-6
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
1630732-B
Compliance level
3 V
rms
3 V/m
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
rms
should be used no closer to any part of the CT515,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance:
d = 1.167√P
d = 1.167√P 80 MHz to 800 MHz
d = 2.333√P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in each
frequency range
Interference may occur in the vicinity of equipment
marked with the following symbol:
Technical Data
b)
.
a)
,
33

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Softcloud pro

Tabla de contenido