SoftCloud® Ace, SoftCloud® Pro
IMPORTANT!
– Wait for the mattress system to gain optimal
pressure.
– Perform a "bottoming out" test after inflating
the mattress following rapid deflation.

5.6 Audible Warning Functions

The audible warning LED flashes, and an audible warning
sounds, to indicate that the control unit or mattress pressure
has failed. The LED will remain illuminated until appropriate
pressure is restored. The audible warning can be silenced by
pressing the audible warning mute button.
The system has four different audible warning signals and
one service indicator, identified by five different symbols.
The symbols are illustrated below.
20
Display
Audible Warning
Signal
Alternating Mode
failure
Low Pressure
Description
Indicates with an audible
sound (press mute button
to silence) and a flashing
LED light when there is no
alternation. The LED light
will change to a fixed light
within 5 min.
Indicates that pressure has
fallen below the tolerance
limit (below setting
pressure –5 mmHg) or
more. This will trigger the
audible warning (press
mute button to silence)
and a flashing LED light.
The LED light will change
to a fixed light within 7
min.
1630732-B