Display
Audible Warning
Signal
Power failure
Initial failure (the
mute button LED
will be constantly
lit)
Service indicator
If audible warning activates and the system fails to
inflate or loses pressure, refer to section "Identifying
and repairing faults".
1630732-B
Description
Indicates no pressure
output due to mains
power failure. This
will trigger the audible
warning (press mute
button to silence) and a
flashing LED light. The LED
light will change to a fixed
light within 3 seconds.
When the power is back
on, the audible warning
and the LED light will
revert to normal mode.
Indicates that the system
did not complete
initialization within 50
minutes, the mute button
LED light will have a fixed
light.
Indicates that service is
needed, the LED light will
have a fixed light. This will
occur when the control
unit has been running for
approximately 9000 hours.

5.7 Transporting a patient on the mattress

1. Before patient transport, press the Mode button and
wait at least 12 minutes for cells to inflate to maximum
pressure.
If a quicker response is required (up to 5-10
minutes), then use the Max Firmness button.
2. Once mattress pressure has reached maximum inflation,
press the Power button to switch off the control unit.
Switch off mains supply and unplug the power cord.
3. Remove the connector from the control unit and allow
air to escape for a few seconds before inserting the
transport cap into the connector air outlets to seal the
system. This release softens the mattress surface for
pressure relief and comfort. Ensure the cap is firmly
secured across both air hose outlets.
CAUTION!
Air pressure is released from all internal static
cells as well as alternating sections.
– Regularly perform a "bottoming out" test to
ensure the patient is appropriately supported.
Usage
21