2.3 Environmental Conditions
Allowable environmental conditions
Temperature range for use: -10 °C to +60°C
Allowable relative humidity 0 % to 90 %, non-condensing
Unallowable environmental conditions
Mechanical vibrations or impacts
Perspiration, urine, fresh water, salt water, acids
Dust, sand, highly hygroscopic particles (e.g. talcum)
2.4 Service Life
This prosthetic component has been tested by the manufacturer for 3 million
load cycles according to ISO 10328. Depending on the patient's activity
level, this corresponds to a service life of 3 to 5 years.
3 Safety
3.1 Explanation of Warning Symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warnings regarding possible technical damage.
NOTICE
3.2 General Safety Instructions
CAUTION
Using the product without following the instructions for use
Deteriorating health condition and damage to the product because of fail
ure to follow the safety information
► Observe the safety information in these instructions for use.
► Make the patient aware of all safety instructions marked with "Inform
the patient".
CAUTION
Excessive strain on the product
Fall due to breakage of load-bearing components
► Use the prosthesis components according to the MOBIS classification
(see section "Area of Application").
► Inform the patient.
14 | Ottobock