7. Squeeze the bottle until a tiny drop forms
at the tip of the bottle. Touch the drop to
the tip of the yellow window of the test
strip. Do not put control solution on top
of the test strip. When you see
you have enough control solution in the test
strip. Wipe the tip of the bottle with a tissue
then cap the bottle tightly.
8. A result appears on the display, along with
the control bottle symbol and a flashing
"L." Do not remove the test strip yet.
Press and release
flash,
result as a Level 1. If you tested the Level 2
control, press and release
time.
once to mark the
a second
21