the screen and may be accompanied by an acoustic signal. The message "outside limits" also ap-
pears on the GLP printout.
6.10 S ensor ID/SN
1. Enter Sensor ID/SN
An alphanumeric sensor ID with up to 12 characters can be entered. The sensor ID will be as-
signed to each calibration and measurement value. This is valuable for tracing back data.
Up to 5 sensor IDs can be entered for each sensor type.
If a new sensor ID is entered, the theoretical calibration slope and offset for this type of electrode
will be loaded. The sensor has to be newly calibrated.
If a sensor ID is entered, which is already in the memory of the meter and has been calibrated be-
fore, the specific calibration data for this sensor ID will be loaded.
When a new ion sensor ID is entered, the electrode type can be selected.
When connecting an ISM
●
Automatically recognize the sensor when it's turned on (alternatively, when pressing READ or
CAL)
●
Load the stored sensor ID, sensor SN and sensor type as well as the latest calibration data of
this sensor
●
Use this calibration for the subsequent measurements
The sensor ID for analog ISM
ISM
sensors.
®
2. Select Sensor ID
Already entered sensor IDs can be selected from a list.
If a sensor ID is selected, which is already in the memory of the meter and has been calibrated be-
fore, the specific calibration data for this sensor ID will be loaded.
You can delete a sensor ID with its calibrations in the calibration data menu.
34 Menus and settings
51710917_V02.2011_BA_SG9_98_EnFrEs.indb 32
sensor to the meter, the meter will:
®
sensors can be changed, However, this is not possible for digital
®
21.04.2011 11:46:16