7.4 I SM data
SevenGo Duo pro™ SG98 and SevenGo pro™ SG9 incorporate Intelligent Sensor Management
(ISM
) technology. This ingenious functionality provides extra security, safety and eliminates mis-
®
takes. The most important features are:
Extra security!
●
After connecting the ISM
and serial number are transferred from the sensor chip to the meter. The data is also printed
on the GLP printout.
●
After calibration of the ISM
to the sensor chip. The most recent data is always stored where it should be – on the sensor
chip!
Extra safety!
After connecting the ISM
These can be reviewed to see the development of the sensor over time. This information provides
an indication if the sensor should be cleaned or renewed.
Eliminate mistakes!
After connecting an ISM
ments.
Additional features are described below.
Initial calibration data for pH (only SG98)
When an ISM
®
trans f erred. The following data is included:
●
Response time
●
Temperature tolerance
●
Membrane resistance
●
Slope (at pH 7) and offset
●
Type (and name) of electrode (for example, InLab
●
Serial number (SN) and ordering (ME) number
●
Production date
Calibration history
The last 5 calibrations data stored in ISM
trans f erred.
Maximum temperature
The maximum temperature that the ISM
i t ored automatically and can be reviewed for the evaluation of the electrode lifetime.
Sensor cap life
The date of first use of the optical DO sensor cap and the expiration date in UTC (universal time
coor d i n at e d) can be reviewed. Once the first measurement is taken with the sensor cap, an internal
clock starts counting down and the cap needs to be replaced after 12 months.
Reset ISM
®
The calibration history in this menu can be deleted. This menu is protected by a deletion PIN. Upon
deliv e ry, the PIN for deletion is set to 000000. Change the PIN to prevent unauthorized access.
38 Data management
51710917_V02.2011_BA_SG9_98_EnFrEs.indb 36
sensor, the sensor is automatically recognized and the sensor ID
®
sensor, the calibration data is automatically stored from the meter
®
sensor, the five most recent calibrations are transferred to the meter.
®
sensor, the last set of calibration data is automatically used for mea s ure-
®
sensor is connected, the initial calibration data in the sensor can be reviewed or
sensor including current calibration can be reviewed or
®
sensor has been exposed to during measurement is mon-
®
Expert Pro ISM)
®
21.04.2011 11:46:18