3. Affix the "Correct posture" label to the third column element.
4. Mark the area shown in the illustration using colored adhesive tape, for
0,5
0,5 m
5. Align the device by turning the foot screws.
Spirit
level
6. Tighten the knurled wheels in the direction of the arrow.
Knurled wheel
WARNING!
Incorrect measurement due to reflections
If there are objects or people in the immediate vicinity of the device,
incorrect measurements will result.
Ensure that there are no objects or people within 0.5 meter of the
front or side of the scale during the measuring process.
Ensure that the device is at least 0.2 meter away from the wall.
Ensure that the patient is not wearing any kind of hair accessory on
top of the head.
example.
The air bubble in the spirit level must be located in the exact center of the
circle.
The foot screws are secured against being adjusted.