Problem
... no beeps can be heard?
... the measured results are not
announced?
... the request to take the printed
ticket is not announced?
... one segment of the multifunctional
display is on either continuously or
not at all?
... the display "StOP" appears?
The display "tEMP" appears.
... the multifunctional display is no
longer reacting to keys being
pressed?
96 •
• Beeps are not activated
- Activate function (see "Activating/deactivating beeps (bEEP)" on page 88)
• Ultrasound measurement is deactivated
- Restart the device
• Volume set to zero
- Increase volume
• Loudspeaker is faulty
- Inform seca Service
• Announcement of measured results is not activated
- Activate function (see "Activating/deactivating announcement of
measured results (reSUL)" on page 82)
• Ultrasound measurement is deactivated
- Restart the device
• Volume set to zero
- Increase volume
• Loudspeaker is faulty
- Inform seca Service
• The function is not activated
- Activate function (see "Activating/deactivating the announcement "Take
your printed ticket" (Print)" on page 83)
• Ultrasound measurement is deactivated
- Restart the device
• Volume set to zero
- Increase volume
• Loudspeaker is faulty
- Inform seca Service
The corresponding point has a fault.
- Inform seca Service
• During weighing:
Maximum load has been exceeded.
- Remove the load from the scale
• When setting up a wireless group:
Wireless channel setup is complete.
- Switch on devices to be integrated in the wireless group (see "Setting up
a wireless group (Lrn)" on page 91)
The ambient temperature of the scale is too high or too low.
- Set up the scale in an ambient temperature between +10 °C and +40 °C
- Wait for around 15 minutes until the scale has adapted to the ambient
temperature.
• Key lock is active.
- Switch off key lock (see "Activating/deactivating key lock" on page 77)
• Device is in an undefined state following implausible input.
- Take power supply unit out of the socket
- Wait approx. 1 minute
- Plug the power supply unit into the socket; the scale and the
multifunctional display will switch on automatically
Cause/solution