Connecting/Positioning the Nasal Cannula

To ensure proper breath detection and oxygen delivery, 
Invacare recommends using a 7 ft (2.1 m) cannula. Tubing 
length not exceeding 25 ft (7.6 m) can be used with 
continuous flow.
NOTE: For this procedure, refer to FIGURE 6.2.
NOTE: The tubing/cannula must be connected to the transportable oxygen 
concentrator oxygen outlet as shown below. 
NOTE: Replace the nasal cannula on a regular basis. Check with your equipment 
provider or physician to determine how often the cannula should be replaced.
NOTE: DO NOT share cannulas between patients.
1.
Connect the nasal cannula to the transportable oxygen concentrator 
oxygen outlet port (DETAIL "A").
2.
Place the cannula over your ears and position the prongs in your nose 
as instructed by your health care provider or cannula manufacturer 
(DETAIL "B").
DETAIL "A"
FIGURE 6.2 Connecting/Positioning the Nasal Cannula

Connecting the Humidifier

(if so prescribed and only with Continuous Flow Mode)
DO NOT use a humidifier while in Pulse Mode.
DO NOT overfill humidifier.
DO NOT reverse the oxygen input and output connections. 
Water from the humidifier bottle will travel through the 
cannula back to the patient.
Part No 1163190