SECTION 7—CLEANING, CARE, AND MAINTENANCE
3.
Lift out the filter.
4.
Use a vacuum cleaner or wash with a mild liquid dish detergent (such 
as Dawn™) and water. Rinse thoroughly. 
5.
Thoroughly dry the filter and inspect for fraying, crumbling, tears and 
holes. Replace filter if damage is found.
6.
Reinstall the air intake filter and snap the filter cover back in place.
Use only Invacare part number 1156863 as the air intake filter for your 
NOTE: 
transportable oxygen concentrator.
Filter Cover
Filter Cover Tab

Cleaning the Cannula

NOTE: To clean the oxygen cannula, follow the instructions provided by the 
manufacturer. If none are provided, follow these steps:
1.
Clean the cannula once a week or as needed.
2.
Wash the cannula in soapy water and rinse it with a solution of ten 
parts water and one part vinegar.
3.
Rinse thoroughly with hot water and hang to dry.

Cleaning the Humidifier

NOTE: To clean the oxygen humidifier, follow the instructions provided by 
manufacturer. If none are provided, follow these steps:
1.
Clean the humidifier every day.
2.
Wash it in soapy water and rinse it with a solution of ten parts water 
and one part vinegar.
3.
Rinse thoroughly with hot water.
Invacare SOLO
2
Filter Cover Tab
FIGURE 7.1 Air Intake Filter
Filter Cover
32
Filter
Part No 1163190