4. Connect the dressing tubing to the canister tubing.
➡
➡
5. Switch Invia Liberty pump
ON by pressing [
If the above check was successful, switch OFF Invia Liberty pump by pressing
[
> 3 seconds ]. If the self-test is not successful, follow instructions shown on
the display or see the Alarms chapter.
For dressing applications, please refer to the instructions for use provided with the Invia dressings. Turn
the Invia Liberty pump on as described in the user modes section.
Invia Liberty pump user modes
Administrative mode
Used by healthcare professional to either set up new patient therapy or to change pump settings such as
pressure, air leakage volume, therapy Constant and Intermittent modes, intermittent times, language and
time zone. You can enter the Administrative mode when first turning the pump on or while therapy is run-
ning.
Patient mode
In Patient mode the pump can be turned on and off, placed in standby mode and an alarm can be
muted. In this mode the canister and tubing can be changed and the pump can be charged. When the
pump is turned off in this mode, last settings are used by default.
18
B.
A.
]
B. "Click" into place.
Canister secures tubing
A. The pump will do a self-test
and start running imme diately.
You should hear the motor
running for a short time until
the pressure is built up. If the
motor keeps running for more
than 30 sec. – check system
and retry.