Maintenance work

Check all parts for deformation and damage and replace them if
deformed or damaged.
Retighten the two screws on the cap (recommended torque:
0.44+0.03 ft lb/60+5 Ncm).
After maintenance work and/or replacement of components
confirm the leak tightness.
Replacing the speech diaphragm (Figure A3 on Page 3)
If fitted, remove the mouthpiece.
Remove the nose cup.
Remove the speech diaphragm using wrench R 26 817 taking
care not to damage the foil on the diaphragm.
Perform visual inspection of speech diaphragm and O-ring and
replace if necessary.
Insert O-ring, then screw in speech diaphragm.
Fasten the nose cup.
Mount the mouthpiece, if necessary.
Use Instructions for NIOSH-approved full facepieces
Replacing the drinking valve
Remove the mouthpiece.
Remove the nose cup.
Remove the covering cap of the drinking valve and remove it
together with the frame.
Loosen the screws (Figure F-1 on Page 3) and remove the rotary
knob (Figure H-6 on Page 3).
Unhook the cap from the visor frame, then remove it from the
connection piece by pulling it away from the drinking valve.
Swivel the drinking valve downwards by 180
the connection piece.
Attach the new drinking valve to the connection piece so that it
points downwards.
Turn the drinking valve into the connection piece so that the safety
lug of the drinking valve grips behind the safety lug of the
connection piece.
Adjust the drinking valve.
Hook the cap into place on the button and place it on the drinking
valve, then hook it into the visor frame.
Pull the covering cap of the drinking valve with the frame up and
away from the drinking valve, then mount the cap.
Mount the rotary knob, then fasten the three screws
(recommended torque: 0.44+0.03 ft lb/60+5 Ncm).
Use Loctite 221 to secure the screw at the rotary knob.
Put the covering cap on the drinking valve.
Fasten the nose cup.
Mount the mouthpiece.