1

8.4 Irrigation level

A. From the Settings mode menu, select the
short press on the knob to change irrigation level.
Note 1
FIG. 4
B. Turn the knob CW or CCW to set up the irrigation level (1).
6 levels of adjustment are possible:
irrigation OFF, 30ml/min, 60ml/min, 90ml/min, 120ml/min,
150ml/min.
Note 3
C. Short press on the knob to exit irrigation level setting.
Irrigation level is saved and the Settings mode menu is
displayed again.

8.5 Contra-angle ratio

A. From the Settings mode menu, select the
short press on the knob to change the contra-angle ratio.
B. Turn the knob CW or CCW to change the contra-angle ratio.
Note 4
C. Short press on the knob to exit the contra-angle ratio setting.
The contra-angle ratio is saved and the Settings mode menu is
displayed again.
FIG. 3
symbol and
symbol and
FIG. 4
NOTES
1
The rotation direction and the irrigation level symbols differ
depending on the actual settings.
2
The Torque value is automatically increased in REVERSE
mode when torquemeter is displayed. The torque value can be in-
creased from 0 to 10 Ncm, see chapter " Reverse torque boost val-
ue" on page 18 to adjust it.
3
When setting the irrigation level to OFF, all dots (1) are dis-
played in black. Irrigation level is off when the irrigation is com-
pletely turned off by means of the footpedal's blue button,
regardless of the active step. In this case, the OFF symbol is dis-
played in Operation mode. The irrigation is considered as a quick
setting and therefore is turned ON when starting again from step
P1.
4
The contra-angle ratio is cyan-colored for direct-drive and
green-colored for reduction gears.
1
17