Step
3
Instructions: Recirculation loop
Step
1
2
3
4
5
6
7
Instructions: Permeate outlets
Step
1
UniFlux 10 Instrucciones de funcionamiento 28993587 AE
Action
In the New Block dialog:
In the Name field, type an appropriate name, for instance
FLOWPATH_CONCENTRATION.
In the Base group box, select Time.
In the Call group box, select From: Block FED_BATCH_CONCENTRATION
and At: 0.00 min.
Click OK to insert the new block.
The new block will be placed inside the Block FED_BATCH_CONCENTRATION
block.
Action
In the Text pane, select the Base Time instruction of the Block
FLOWPATH_CONCENTRATION block.
In the Instruction box:
Set Breakpoint to 0.00 min.
In the Instructions group box, select Recirc.
Set Feed to Open, click Insert.
Set Feed_Drain to Closed, click Insert.
Set IntegrityTestValve to Closed, click Insert.
Set Retentate_Control_Valve_PCV341 to 100%, click Insert.
Set Retentate to Retentate, click Insert.
Action
In the Text pane, select the Base Time instruction of the Block
FLOWPATH_CONCENTRATION block.
A Create a method with UNICORN 5
A.2 An example of using the Method Editor
A.2.5 Block 3: Fed-batch concentration
207