B Create a method with UNICORN 6 and above
B.1 The Method Editor
B.1.3 Other Method Editor features
Totalizers
A totalizer monitors the accumulated volumes passing through the flow meters. For
each totalizer there is a watch instruction allowing easy programming of desired actions
triggered by specific events.
Follow the instruction below to define an action based on the totalizer, for instance for
the permeate volume.
Step
1
2
The allowed noise, or error, for a totalizer can be defined using the appropriate
FT14VTot_WatchPar instruction (under Specials).
226
Action
Reset the totalizer to 0 liters by inserting instruction FT142VTot_Reset (under
Permeate).
Insert watch instruction (under Watch) and set values for Signal (FT142Vtot),
Test (e.g., Greater_Than), Value (volume in liters) and Action (e.g., End block).
UniFlux 10 Instrucciones de funcionamiento 28993587 AE