Step
3
4
Instruction: Setting Delta P
watch
Step
1
2
3
4
Set block end time
Change the block end time to 10 minutes to give Delta P some time to stabilize.
Step
1
2
3
UniFlux 10 Instrucciones de funcionamiento 28993587 AE
Action
In the Parameters for field:
Set BaseLine to for instance 0.1 bar.
Set MinPeak to 0.00 bar.
Click Insert.
Action
In the Text Instructions tab, select the Base: SameAsMain instruction of the
Block: STABLE_DELTA_P block.
In the Instruction Box:
Set Breakpoint to 0.00 min.
In the Instructions: list, select Watch and Watch.
In the Parameters for field:
Set Signal to DeltaP.
Set Test to Stable signal.
Set Minutes to 1.0.
Set Action to End block.
Click Insert.
Action
In the Text Instructions tab, select the End_Block instruction of the Block:
STABLE_DELTA_P block.
In the Instruction Box, set Breakpoint to 10.00 min.
Click Change.
B Create a method with UNICORN 6 and above
B.2 An example of using the Method Editor
B.2.5 Phase 3: Fed Batch Concentration
251