GE Uniflux 10 Instrucciones De Funcionamiento página 204

Tabla de contenido

Publicidad

A Create a method with UNICORN 5
A.2 An example of using the Method Editor
A.2.4 Block 2: Product transfer
Instruction: Tank level watch
Step
1
2
End block: Allow the tank to be
filled
Step
1
2
204
Action
In the Text pane, select the Base Time instruction of the Block
PRODUCT_TRANSFER_VOLUME block.
In the Instruction box:
Set Breakpoint to 0.00 min.
In the Instructions group box, select Watch and WatchTankLevel.
Set:
1 Test to Greater_Than.
2 Value to a number in liters given by the tank specification or the
application requirements.
3 Action to END_BLOCK.
Click the Insert button.
Action
In the Text pane, select the End_Block instruction of the Block
PRODUCT_TRANSFER_VOLUME block.
In the Instruction box:
Set Breakpoint to a value high enough for the tank to reach the volume
set for WatchTankLevel (in the previous step).
Click the Change button.
UniFlux 10 Instrucciones de funcionamiento 28993587 AE

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido