Setting up the SIRIUS Multigas Detector

5.3 Instrument Alarm Bypass Options

The SIRIUS Multigas Detector (with software version 1.1 or higher) is equipped with a
feature to disable or silence the visual, backlight, and audible options. If any of these
options are disabled during instrument start up, the SIRIUS Detector displays:
"VISUAL OFF" if the red LEDs are disabled
"AUDIBLE OFF" if the audible buzzer is disabled
"BACKLITE OFF" if the backlight is disabled
"BACKLITE TIME"
If the visual or audible options are disabled, "ALARM OFF" flashes on the LCD during
Normal Measure mode.
1. Instrument Options Setup
Safe LED ON/OFF
Operating beep ON/OFF
STEL/TWA ON/OFF
2. CAL Lockout Enable
To disable calibration in the user mode, set CAL LOCK MODE to ON
When ON, calibration is only accessible via the Setup mode and password
(if enabled)
3. CAL DUE Alert
To disable CAL Due messages, turn this feature OFF.
When ON, the number of days (1 to 180) between calibrations can be set.
User will be required to acknowledge an overdue calibration at turn ON.
4. Warm Up Info
Turning this selection OFF causes the instrument NOT to display alarm set
points at power-ON
Time (if data logging option installed)
Date (if data logging option installed)
5. LEL/CH
Setup
4
Sensor ON/OFF (turns the sensor ON or OFF)
Display Combustible Gas Type?
Methane
Pentane
Hydrogen
Propane
LEL or CH
Low Alarm (sets the low combustible alarm)
High Alarm (sets the high combustible alarm)
Cal Gas (sets the expected combustible calibration gas)
30
mode displays % LEL (for any gas) or % CH
4
(for Methane only)
4