9
3. After finishing Year setting, you can press "TIME" button again.
Now you can adjust Month/Date/Hour and Minute digits blink as you desire.
TROUBLESHOOTING
Phenomenon
Nothing
is
displayed
when
you
push
the
Power/Start button
"Err P" shown on
LCD panel
"Err 1" shown on
LCD panel
"Err 2" shown on
LCD panel
"Err 3" shown on
LCD panel
lf correct measurement is impossible even after checking the above-mentioned points,
consult at the store where you have purchased the unit or the nearest dealer without touching
the internal mechanism.
In some very rare cases, there may be error due to the physical condition of the person. In
such cases, please consult your doctor.
SPECIFICATIONS
Model: Desktop Arm Blood Pressure Monitor
Measurement: Oscillometric
Measurement Range: Pressure: 20 - 280 mmHg,
Accuracy: Pressure: within 1 3 mmHg,
Pulse rate: within 1 5% of reading
(subject to the temperature, humidity listed below)
Inflation: Automatic inflation via pump
Pressure Detection: Semi - conductor
Problem
Are
the
batteries
exhausted?
Are the polarities of
batteries in the wrong
position?
Fail to inflate
Deflate too fast
1.Movements occur on
the arm
2. Interference signal
around the BPM
Incorrect
measured
result
Pulse Rate: 40 - 195 beats I min
ENGLISH
Solution
Replace the batteries with new ones
1. Please check if the connection on main
unit and cuff are loosen. Then connect
them.
2. Please check if air tube is broken and
change it.
Please send it to service station.
1. Keep still after you push Power/Start
button.
2. Please remove interfering sources such
as magnetic material or cell phone.
1. You might wear thick jacket and your
arm fails to detect pressure values.
Please take of your jacket and press BPM
again.
2. You might use it incorrectly such as
measure pressure value on child