Chapter 15
TROUBLESHOOTING
Abnormal Phenomenons
BP measurement values too high
or too low
No pressure
Cuff deflates in short time
It can not carry on measurement when press the measurement button
Power off suddenly when inflating
Hold the on/off button but can not
start the device
Cuff inflation start before press the measurement button or never stop
inflating when measuring
Cuff never deflation
Causes
Cuff is not connected correctly
Move limbs when measuring
Cuff leakage
The cuff windpipe is not correctly
connected with cuff
Cuff is not inflated
Loose cuff
No use for a long time, the power of
batteries can be exhausted owing to
the changed temperature
Power of batteries can be exhausted
The battery polarities is reversed
ENGLISH
Solutions
Correctly connect cuff (refer to Chapter 9)
Keep quiet and restart a measurement
Buy a new cuff
Correctly connect
Stop using the device and contact us
Correctly apply cuff
Switch on the power once again and
restart a measurement
Replace all four batteries with new ones.
Replace all four batteries with new ones
Check the battery installation for proper
placement of the battery polarities.
Pull out the cuff to deflate. Stop using the
device and contact us
Pull out the cuff to deflate. Stop using the
device and contact us