Troubleshooting - Gima 32926 Manual Del Usuario

Ocultar thumbs Ver también para 32926:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Storage
• If the device is not to be used for a long time, please remove the batteries from the device (leaking of
battery acid can cause the device to malfunction).
• Always store the unit in the storage case after use.
• Do not place the device directly under sunlight, in high temperature, or in humid or dusty places.
• Do not store the device in extremely low (less than -13°F/-25°C) and high (more than 158 °F/70 °C) temper-
ature, nor in a place its humidity exceeds 93% R.H.

TROUBLESHOOTING

SYMBOLS/SYMPTOMS
No power-on by pressing
button.
Measuring Error Symbol appears
when blood pressure value dis-
played is excessively low or high.
Measuring Error Symbol
Measuring Error Symbol
Measuring Error Symbol
Note: If "EP" appears on the display, just return the device to your local distributor or importer.
ENGLISH
CONDITIONS/CAUSES
Worn-out batteries.
Battery polarities have been
positioned incorrectly.
Cuff has been placed incorrectly.
Did you talk or move during
measurement?
Shaking of the wrist with the
cuff on.
Air circuit abnormality.
Cuff tube may not be plugged into
monitor correctly.
Inflation pressure exceeding
300 mmHg.
Error determining measurement
data.
INDICATION/CORRECTION
Replace them with 2 new AAA
alkaline batteries.
Re-insert the batteries in the
correct positions.
Wrap the cuff properly so that it is
positioned correctly.
Measure again. Keep wrist steady
during measurement.
Check cuff connection.
Measure again.
Switch the unit off, then measure
again.
Measure again.
36