2 Safety information
2.1 Basic safety information
8
2
Safety information
2.1
Basic safety information
2.1.1
Prerequisites
NOTICE
Fragment everything excluding APOLLO DI
Important information on building installation
The building installation must be performed by a qualified expert in
compliance with the national regulations. DIN VDE 0100-710 applies in
Germany.
NOTICE
Restrictions regarding installation site
The system is not intended for operation in areas subject to explosion
hazards.
NOTICE
Do not damage the unit!
The unit can be damaged if opened improperly.
It is expressly prohibited to open the unit with tools!
2.1.2
Operating
The suction system must only be operated with an integrated filter bag
and HEPA filter. Only the approved filter bag and HEPA filter may be used
(see "Spare parts and consumables [ → 24]").
2.1.3
Maintenance and repair
As manufacturers of dental instruments and laboratory equipment, we
can assume responsibility for the safety properties of the unit only if the
following points are observed:
● The maintenance and repair of this unit may be performed only by
Sirona or by agencies authorized by Sirona.
● Components which have failed and influence the safety of the unit
must be replaced with original (OEM) spare parts.
Please request a certificate whenever you have such work performed. It
should include:
● The type and scope of work.
● Any changes made in the rated parameters or working range.
● Date, name of company and signature.
2.1.4
Modifications to the product
Modifications to this product which may affect the safety of the operator,
patients or third parties are prohibited by law!
Operating Instructions Suction devices for inLab MC X5 and