16
Instructions for Use
Use

Controller Controls, Indicators and Connectors

The following paragraph illustrates the Controller control panel and
interconnection panel. More details are contained in the appendix
"Technical Information".
Figure 2
1
Keyboard push-button for LOW SPEED mode selection. It is active only when the front
panel operation has been selected. Pressed once, the pump runs at about 2/3 of the
nominal speed. To unselect the mode, press the push-button again.
2
Keyboard push-button for START, STOP, RESET mode selection. It is active only when the
front panel operation has been selected. By pressing once the starting phase begins; if
pressed again it stops the pump. If the pump has been stopped automatically by a fault, this
push-button must be pressed once to reset the controller and a second time to restart the
pump.
3
Keyboard push-button to recall on the display the cycle number, cycle time and pumplife.
4
Keyboard push-button to recall on the display the upper bearing temperature, lower bearing
temperature, pump power and rotational speed. It is always active regardless of the
operating mode selected. Push-buttons 3 and 4, if pressed together for at least 2 seconds
put the controller in a routine where it is possible to program some operation parameters.
5
LCD back-lighted alphanumeric display: dot matrix 2 lines x 16 characters.
230/306
Controller 969-9462, 969-9562 front panel