Technical Information
Use
Figure 20
where: XXXXXX = Argon or Air
If Argon is selected the maximum power delivered to the eletrical
motor allows the pumps to work properly with heavy gases such
Argon, Silane and Arsine). If Air is selected the maximum power
delivered to the electrical motor allows the pump to work properly
with light gases such as Nitrogen, Helium and Hydrogen. The
controller is factory set to Argon.
272/306
Enter the selection by pressing the PUMP CURRENT push-button,
and the display shows:
P
R
O
C
X
X
Turbo-V 2000HT Controller User Manual / 87-900-913-01
E
S
S
X
X
X
G
A
S
?
(O)