MESSAGE
Turbo-V 2000HT Controller User Manual / 87-900-913-01
DESCRIPTION
In normal operation (after the starting
phase) the current drawn by the pump
is higher than programmed (8 A).
After the starting phase the output
connection is shorted (output current
higher than 38 A).
The pump is stopped by an emergency
stop signal provided via a remote
contact.
The controller has received a random
signal or the line voltage is not
compatible with the setting of the
internal voltage switch.
Instructions for Use
Error Messages
REPAIR ACTION
Check that the pump rotor is free
to rotate.
Press the START push-button
twice to start the pump.
Check connections and shortages
between pump and controller.
Press the START push-button
twice to start the pump.
Remove the controller power
cable and check the emergency
condition. Then reconnect the
power cable and
press the START push-button
twice
to start the pump.
Press the START push-button
twice to start the pump.
Should the message still be
present, call the Agilent service
(O)
16
235/306