Input Pressure And Equipment; Pneumatic Connection - GE PACE1000 Instrucciones De Seguridad Y Guía Del Usuario

Equipo automatizado de calibración de presión
Ocultar thumbs Ver también para PACE1000:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 75
2.4 Input Pressure (Figure 3)
1. Make sure the user systems can be isolated and vented.
2. Connect the Unit Under Test (UUT) to the output connection port.
Note: For instruments with NPT connections, use applicable bonded sealing (2) as shown
in figure below.
1) Connector
3 Installation
The instrument connects to the Unit Under Test.

3.1 Input Pressure and Equipment

The pressure should not exceed 1.25 x full-scale or MWP stated on the rear panel of the
instrument.
To protect the instrument from over-pressure a suitable protection device (such as a relief valve or
bursting disc) must be fitted to prevent over pressurization.

3.2 Pneumatic Connection

WARNING
PRESSURE RANGES > 210 bar (3000 psi) ARE ONLY RATED FOR HYDRAULIC USE.
CAUTIONS
Do not exceed the maximum pressures stated in the appropriate Component
Manual for the unit under test.
Reduce pressure at a controlled rate when venting to atmosphere.
Carefully de-pressurize all pipes to atmospheric pressure before disconnecting and
connecting to the unit under test.
[EN] English
Figure 3 Pneumatic Connections
2
1
2) Bonded seal
6
K0467 Revision A

Publicidad

Tabla de contenido
loading

Tabla de contenido