cAring for your igo system

SYMPTOM
Red Service required light is flashing. Flow
Indicator light is flashing. Audible Alert is
sounding. Unit is operating.
Unit does not operate when using with
charged battery. Power Fail alarm activates.
DC Adapter is attached. Unit runs from
Battery or does not run. The External
Power Present Light not illuminated.
Lowest Battery Status Light is Yellow. Unit
beeps once.
Battery Status Lights never indicate fully
charged.
Yellow Battery Status light is flashing.
Audible Alert sounds intermittently when
operating from DC Adapter.
If any other problems occur with your iGo.
CaRInG foR YoUR DevIlbIss IGo PoRTable oXYGen sYsTeM
Before attempting any cleaning procedures, turn the unit Off�
NOTE
-Use no lubricants, oils, or grease.

Air Filter

The air filter should be cleaned at least once a week. To clean, follow these steps:
1.
Remove the air filter.
2.
Wash in a solution of warm water and dishwashing detergent.
3.
Rinse thoroughly with warm tap water and towel dry. The filter should be completely dry before reinstalling.
CAUTION-To prevent product damage, do not attempt to operate the unit without the air filter or while the filter is still damp.
AC Adapter and Power Cord/DC Adapter
The adapters and power cords should be cleaned as needed. To clean, follow these steps:
1.
Disconnect cords from the iGo before cleaning.
2.
Clean the cords by using a damp cloth with a mild household cleaner and wipe dry.
Cannula and Tubing
Clean and replace the cannula and tubing as instructed by your DeVilbiss provider.
POSSIBLE CAUSE
1. Blocked or defective cannula or oxygen
tubing
1. Battery is overheating.
1. Poor connection to DC power source.
2. Vehicle power source (vehicle accessory
connector) dropped too low for the DC
Adapter.
3. Fuse in vehicle is blown.
1. Battery needs to be charged.
1. Battery needs to be conditioned.
2. Defective battery.
1. Unit is equipped with built-in battery test and
the battery pack is being checked.
1. Vehicle not running.
2. Poor connection to DC power source.
3. Vehicle electrical system overloaded or
defective.
� WArning
REMEDY
1. Detach cannula. If proper flow is restored,
clean or replace if necessary. Disconnect the
oxygen tubing at the oxygen outlet. If proper
flow is restored, check oxygen tubing for
obstructions or kinks. Replace if necessary.
1. Allow battery to cool.
1. Ensure the DC accessory outlet is clean and a
good connection can be made. Insert the
other end of the DC Adapter into the
vehicle's DC accessory outlet.
2. If the vehicle power source drops too low for
the DC Adapter, the iGo will revert to
Battery operation (if present) or will cease
operation until power is restored.
3. Check the fuse and replace if necessary.
1. Plug unit into AC power or replace
discharged battery with charged battery.
2. Plug unit into DC accessory outlet to operate
unit (battery will not recharge on DC power).
1. Fully discharge battery then recharge.
2. Contact your DeVilbiss provider.
1. If light flashes longer than 5 minutes, battery
is defective. Contact your DeVilbiss provider
for a replacement.
1. Start vehicle.
2. Ensure the DC accessory power port outlet
is clean and a good connection can be made.
Insert the DC Adapter into the vehicle's DC
accessory outlet.
3. Have a qualified auto technician check
electrical system while iGo is attached.
1. Turn your unit Off and switch to your reserve
oxygen system. Contact your DeVilbiss
provider immediately.
EN
A-306
13