Before Operating Your Igo; Selecting The Igo Power Source; Ac Operation; Dc Operation - DeVilbiss Healthcare iGo 306DS Manual Del Usuario

Sistema de oxígeno portátil
Ocultar thumbs Ver también para iGo 306DS:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 16
EN
6.
Position your unit near an electrical outlet at least 6 inches (16 cm) from walls, draperies, or any other objects that might prevent the proper
flow of air in and out of your iGo. The iGo should be located so as to avoid pollutants or fumes, and placed in a well -ventilated place so that
the air inlet and exhaust are not blocked. Do not cover unit with a blanket, towel, quilt, or other covering, as the unit may overheat.
NOTE–
Do not connect the iGo to an extension cord or to an electrical outlet controlled by a wall switch. No other appliances should be
plugged into the wall outlet.

Before Operating Your iGo

1.
Always check to see that the Air Filter is clean. Proper cleaning of this filter is discussed in the Caring for Your iGo System section.
2.
Check to be sure that the Exhaust Vents are not blocked.
3.
Attach the appropriate accessories to the oxygen outlet.
Oxygen Tubing Connection:
a.
Attach the tubing to the oxygen outlet. Make sure it is securely attached.
Oxygen Tubing Connection with Humidification:
NOTE–
A humidifier can be used in Continuous Flow mode ONLY. Do not use a humidifier when in PulseDose mode
a.
Attach the elbow humidifier adapter directly to the humidifier bottle. Attach oxygen tubing (50' max) to the to the elbow humidifier
adapter and oxygen outlet. Make sure it is securely tightened. Place the humidifier bottle on the humidifier stand and secure the stand on
a sturdy, flat surface near the patient where it will be safe and will not be knocked over. The humidifier bottle must remain upright to
prevent water from entering the cannula.
4.
Your physician has prescribed a nasal cannula. In most cases, they are already attached to the oxygen tubing If not, follow the manufacturer's
instructions for attachment. Attach the cannula tubing (7' max) directly to the humidifier bottle.
NOTE–
Use only standard nasal cannula when in PulseDose mode; do not use pediatric (low-flow) nasal cannula with PulseDose
delivery. Any nasal cannula can be used with Continuous Flow mode.
5.
For portable use, ensure fully charged battery is installed and necessary accessories (e.g. AC or DC cords/adapters) are packed.

Selecting the iGo Power Source

The iGo is a lightweight, portable oxygen concentrator capable of being operated directly from three different power sources: (1) AC Power (2)
DC Power, or (3) Rechargeable Battery.
Do not expose any power cords/adapters to water or other liquid spills or drips� Electric shock or serious injury may result�
Improper use of the power cords, plugs, and/or adapters can cause a burn, fire, or electric shock hazard� Do not use any
power cord/adapter that is damaged�
Do not operate device or accessories in standing water and do not submerse or expose to water� The iGo enclosure does not
provide protection against the harmful effects of liquid ingress� Electric shock or damage to the unit may result�

AC Operation

The iGo includes a universal AC Adapter which attaches to the AC Power Cord for use at home or where standard AC power is available. To use AC
power:
1.
Connect the AC Adapter by aligning the arrow on the plug locking ring with the white dot on the receptacle of the unit. Push until the locking
ring twists slightly counterclockwise, then snaps back into alignment with the dot on the receptacle. To disconnect: grasp the plug, twist
counterclockwise, and pull. When properly connected, a green indicator light on the AC Adapter will light and the External Power Present
light will appear on the control panel.
2.
Connect the AC Power Cord to the AC Adapter and to a grounded power outlet. Do not connect the iGo to an extension cord or to an
electrical outlet controlled by a wall switch. No other appliances should be plugged into the wall outlet.
NOTE–
Use with grounded, 3-pin receptacle. DO NOT remove ground pin from cord. Use only with DeVilbiss approved cords.
NOTE–
The AC Power Cord used with the iGo must meet the requirements of the country where the product is used.

DC Operation

Your iGo includes a DC Adapter that allows the system to operate from DC accessory power port outlets such as those found in motor vehicles.
1.
Determine the Fuse rating in your motor vehicle; the rating must be 15 AMP or greater —-consult the Operator's Manual for your vehicle or
look at your fuse panel in the vehicle.
Start your vehicle. NOTE–
2.
the vehicle while the DC Adapter is connected to the vehicle. This may drain the vehicle's battery.
3.
Insert the DC Adapter into the Power Input on the side of the iGo. Connect the adapter by aligning the arrow on the plug locking ring with
the white dot on the receptacle of the unit. Push until the locking ring twists slightly counterclockwise, then snaps back into alignment with
8
A-306
� WArning
DO NOT leave the iGo or DC Adapter plugged into the vehicle without the engine running or attempt to start
set up
: